Abstract
More widespread use of positron emission tomography (PET) imaging is limited by its high cost and radiation dose. Reductions in PET scan time or radiotracer dosage typically degrade diagnostic image quality (DIQ). Deep-learning-based reconstruction may improve DIQ, but such methods have not been clinically evaluated in a realistic multicenter, multivendor environment. In this study, we evaluated the performance and generalizability of a deep-learning-based image-quality enhancement algorithm applied to fourfold reduced-count whole-body PET in a realistic clinical oncologic imaging environment with multiple blinded readers, institutions, and scanner types. We demonstrate that the low-count-enhanced scans were noninferior to the standard scans in DIQ (p < 0.05) and overall diagnostic confidence (p < 0.001) independent of the underlying PET scanner used. Lesion detection for the low-count-enhanced scans had a high patient-level sensitivity of 0.94 (0.83–0.99) and specificity of 0.98 (0.95–0.99). Interscan kappa agreement of 0.85 was comparable to intrareader (0.88) and pairwise inter-reader agreements (maximum of 0.72). SUV quantification was comparable in the reference regions and lesions (lowest p-value=0.59) and had high correlation (lowest CCC = 0.94). Thus, we demonstrated that deep learning can be used to restore diagnostic image quality and maintain SUV accuracy for fourfold reduced-count PET scans, with interscan variations in lesion depiction, lower than intra- and interreader variations. This method generalized to an external validation set of clinical patients from multiple institutions and scanner types. Overall, this method may enable either dose or exam-duration reduction, increasing safety and lowering the cost of PET imaging.
Similar content being viewed by others
Introduction
Positron emission tomography (PET) imaging is used for a wide range of clinical indications, including detecting, staging, and restaging tumors, dementia, and epilepsy, despite its relatively high cost and its use of radioactivity1. PET images are obtained by injecting patients with a standardized dose of a radiopharmaceutical (e.g., 18F-fluorodeoxyglucose [FDG]). The image quality is proportional to the number of coincidence events in the PET detector following radiopharmaceutical positron annihilation. Using the PET images, a quantitative standardized uptake value (SUV) can be calculated by normalizing the radiotracer uptake with its dosage and patient weight or lean body mass2,3. The maximum SUV (SUVmax) is widely used as a semiquantitative measure of the tumor glucose metabolism4.
PET is commonly used in conjunction with computed tomography (CT) imaging to provide attenuation correction and anatomic localization. However, whole-body PET/CT scans can require upward of 30 min of table time, depending on the field of view and patient’s height. This can cause patient discomfort and anxiety, and may potentially lead to motion artifacts that degrade image quality and SUV quantitation5,6. Additionally, exposure to radiation during diagnostic imaging can increase cancer risk, especially in pediatric populations7,8, and raises concerns about exposure to technologists9. While such concerns make scanning with reduced bed times or lower radiotracer dose appealing, both actions lower detection of PET annihilation events, which reduces image quality and SUV accuracy. Consequently, reducing PET scan durations or using lower radiotracer dose without reducing the diagnostic image quality (DIQ) or biasing SUV measurements would be clinically valuable and would enable a higher throughput of patients for diagnostic imaging10. Recently, with the looming burden of COVID-19, which necessitates additional cleaning and sanitization of PET scanners, increased scanning efficiency will limit the bottlenecks in PET imaging for maintaining an adequate patient throughput.
Advances in deep learning and convolutional neural networks (CNNs) have presented an exciting opportunity in medicine, primarily for solving a variety of image-classification problems in radiology, pathology, and dermatology11,12,13,14. These methods have also been used to improve the quality of diagnostic radiology images for several imaging modalities15,16, and specifically, for synthesizing high-quality PET images from input images acquired either with a low radiotracer dosage or acquired over a shorter duration17,18. A handful of recent studies have demonstrated the denoising capabilities of CNNs to enhance PET image quality in small patient cohorts19,20,21. While most of the preliminary studies have focused on enhancing image quality, they have not been deployed in a realistic clinical setting and the impact on quantitative SUV measurements has not been robustly validated across vendors and reconstruction algorithms. Unlike other imaging modalities, to our knowledge, there have been no studies evaluating the clinical utility of low-count-enhanced PET scans22,23.
There are also many PET imaging vendors, each with unique hardware and software that can considerably affect overall image quality24. CNNs trained on a specific domain of inputs have consistently demonstrated limited generalizability when data from a different domain are presented to the network25. In medical imaging, it is becoming increasingly well-known that models developed in specific, narrow populations fail to generalize well to data from different institutions and scanner types26,27. Thus, a key question that has not been addressed with current low-count PET CNNs is that of model generalizability, i.e., how well can a model trained on one subset of data generalize to new unseen patients from different institutions and different scanners28.
To overcome the aforementioned challenges, in this study, we utilize deep learning to enhance the quality and maintain quantitative SUV accuracy of FDG PET scans acquired or simulated with 4x lower counts (i.e., equivalent to either four-fold faster or fourfold reduced dose). We evaluate both qualitative and quantitative performance in an external validation cohort drawn from cancer patients at multiple institutions scanned with a variety of PET scanner devices under typical clinical conditions. We hypothesize that deep-learning-enhanced fourfold reduced-count PET is noninferior to current standard-dose FDG PET for clinical evaluation purposes and is equivalent for SUV-based quantification of tumor burden.
Results
Low-count-enhanced images
Representative whole-body PET images for the 25% low-count images, the 25% low-count-enhanced images using the CNN, and the corresponding 100% dose-standard images for all three scanners used in this study are shown in Figs. 1–3. Figure 1 shows three different patients all with a BMI of less than 30, with oropharyngeal cancer (Fig. 1a), lymphoma (Fig. 1b), and lung cancer (Fig. 1c). Similar images are shown for subjects with BMI over 30, with colon cancer (Fig. 2a), lymphoma (Fig. 2b), and metastatic carcinoma of the head and neck (Fig. 2c). The effectiveness of the deep-learning enhancement technique in improving image quality and lesion conspicuity compared with the low-count scans is shown in Fig. 3. Images from an example subject with a stump were also successfully enhanced using the enhancement model without creating any artifacts, despite the images being out-of-distribution compared with the training data (Supplementary Fig. 1).
Image quality assessment
The DIQ ratings for the standard and low-count deep-learning enhanced scans were 4.0 ± 0.7, and 3.7 ± 0.7 respectively, while the ODC ratings were 4.5 ± 0.6 and 4.2 ± 0.8, respectively. DIQ and ODC values assigned by all three readers (Fig. 4) were comparable between the standard and low-count-enhanced PET scans, and were consistently higher than the minimum diagnostically acceptable quality (score of 3). Both the overall DIQ (p = 0.77) and ODC (p = 0.44) scores did not significantly vary as a function of the scanner used, but they did vary as a function of reader for both DIQ and ODC (both p < 0.01). The reader-based overall score heterogeneity was a random variable and was accounted for in the noninferiority analysis using the RMSE of the model. The interaction term between the imaging method (low-count-only and low-count-enhanced scans) and the scanners was nonsignificant for both DIQ (p = 0.59) and ODC (p = 0.28), showing that the difference in scores and image-quality enhancement between the low-count-only and low-count-enhanced images did not depend on the underlying scanner used.
The point estimate for differences between the standard and low-count-enhanced methods and their 95% confidence intervals (CI) for DIQ was 0.35 (95% CI of 0.21-0.49) and for ODC was 0.25 (95% CI of 0.14–0.36), with the respective p-values of 0.02 and <0.001 (Supplementary Figure 2). The results demonstrated the noninferiority for both DIQ and ODC with the proposed technique. The standard deviations for ODC and DIQ differences between the two imaging methods were 0.67 and 0.81, respectively, both of which were within the maximum tolerable standard deviation for a sample size of 50 patients29.
All standard PET scans maintained a DIQ and ODC of three (clinically acceptable) or higher. In total, 8/180 (3.9%) and 2/180 (1.1%) of the reads of the low-count-enhanced scans were scored as 2 (poor) for DIQ and ODC, respectively. None were scored as 1 (nondiagnostic). On an individual reader basis, this was distributed in the following manner: DIQ (two cases for reader 1, five cases for reader 2, and one case for reader 3) and ODC (one case each for readers 1 and 3).
Repeatability assessment
ICC values for interreader repeatability of DIQ and ODC were 0.25 and 0.33, respectively. The interreader rating confusion matrices (Supplementary Tables 1 and 2) demonstrated that although the readers had different proclivities in providing scores of 3, 4, and 5 (Supplementary Fig. 3), all readers consistently graded the standard scans as similar to the low-count-enhanced ones. For the 10 cases that were read twice, there were no significant intrareader variations for any reader for both DIQ and ODC (DIQ p-values of 0.62, 0.12, and 0.02 and ODC p-values of 0.71, 0.05, and 0.32 for readers 1–3, respectively, with a Holm–Bonferroni-adjusted critical p-value of 0.017). Confusion matrices for intrareader repeatability are provided in Supplementary Tables 3 and 4.
Lesion detection
A combined total of 491 hypermetabolic lesions were detected in 92 total standard scans read by the three readers. There were no hypermetabolic lesions detected in the brain. The number of hypermetabolic lesions detected on the standard and low-count-enhanced scans were not statistically different (p = 0.30) and there was a strong correlation in the number of lesions detected on the two scans (CCC 0.88, more detail in Table 1). There were no systematic biases in up-staging or downstaging subjects based on the number of lesions detected on the low-count-enhanced scans compared with the standard scans for all organs evaluated (Table 1). An example case of a subject receiving a higher DIQ score on the LCE scan than the standard scan in the case of a lung nodule is shown in Supplementary Fig. 4.
The overall patient-level sensitivity and specificity for detecting lesions on the low-count-enhanced scan compared with the standard scan was 0.94 (0.83–0.99) and 0.98 (0.95–0.99) (Table 1). There were no significant differences in the total and organwise number of hypermetabolic lesions detected between the repeated reads pooled across all readers (lowest p-value = 0.48), indicating high repeatability of the readings. Example images from a subject with subtle liver lesions and a noisy low-count scan depicted that the low-count-enhancement technique successfully maintained high lesion conspicuity, despite starting with noisy images (Supplementary Fig. 5). Overall, there were eight instances of false positives or false negatives across 50 patients and six tissue types as assessed by the three readers. The distribution of these deviations did not follow any specific pattern as a function of the tissue of institution—bone (one false positive from Institution B), lymph nodes (one false negative from Institution C), lung (one false positive and false negative each from Institution A), spleen (two false positives from Institution B), and muscle (one false positive and false negative each from Institution C).
The Cohen’s kappa value of 0.85 (95% confidence interval 0.81–0.90) for inter-scan agreement between the lesions detected by the standard and low-count-enhanced scan was nearly identical to the kappa of 0.88 (0.82–0.95) for intrareader agreement between repeated readings for the same patient by the same reader. Similarly, the agreement between the low-count-enhanced and standard scans was higher than that for inter-reader agreement for all pairs of readers: Cohen’s kappa for readers 1 and 2 = 0.58 (0.50–0.66); readers 1 and 3 = 0.69 (0.62–0.76); readers 2 and 3 = 0.72 (0.65–0.79) (Table 2).
SUV equivalence
In total, 99 hypermetabolic lesions were identified on the standard PET scans by the fourth reader. CCC and Bland–Altman plots for SUVmean of aortic blood pool, liver, and right gluteus muscle reference regions and the SUVmax of the lesions showed minimal variation among the standard and low-count-enhanced scans (Fig. 5). The mean difference of the SUVs was approximately zero (ranging from −0.05 to 0.01 for the different regions), with tight 95% confidence limits of agreement (0.10–0.20 for the SUVmean of the reference regions, and 1.8 for the SUVmax of the lesions) (Table 3). Correlation coefficients were very high between the two scans: aortic blood pool (CCC = 0.94), liver (CCC = 0.98), gluteus muscle (CCC = 0.96), and lesions (CCC = 0.99). No statistically significant differences were found for SUV values for the aortic blood pool (p = 0.91), liver (p = 0.80), gluteus muscle (p = 0.59), and lesions (p = 0.79). Using an SUVmax threshold of 2.5, the sensitivity and specificity of lesion detection on the low-count-enhanced scans was 0.98 (0.93–1.00) and 1.0 (1.0–1.0). For the determination of a lesion as a function of its SUV, two false negatives occurred when the SUVmax values on the standard scans were 2.8 and 2.7, but on the low-count-enhanced scans were 2.4 and 2.1, respectively. Both cases arose from different subjects from Institution C.
Discussion
In this study, we demonstrated noninferiority of deep learning to enhance noisy PET images acquired with 4-times fewer counts for clinical purposes in a blinded, multicenter study. Quantitative accuracy for SUV measurements and depiction of hypermetabolic lesions were also maintained. Compared with other low-count PET-enhancement methods using CNNs, the findings in this study were demonstrated in a clinically relevant, external validation cohort that consisted of subjects with varying pathological findings that were assessed by board-certified nuclear medicine physicians. Moreover, the patients were scanned at different institutions on different vendor hardware that used different image-reconstruction algorithms, none of which was used to train the CNN model. We therefore conclude that the low-count-enhancement algorithm enabled the acquisition of PET data four-times faster or at fourfold reduced dose while providing similar diagnostic information as a standard PET examination in a generalizable manner across PET vendors and scanner models.
Since faster and lower-dose scans lead to similar image quality in this regime, the method demonstrated could be used either to increase efficiency or to reduce dose30,31. Increased efficiency is beneficial, given the potential to reduce the cost of PET and enhance patient throughput by scanning more patients on a single scanner. Decreased bed times for PET imaging would also allow for additional duration for sanitization protocols in trying to mitigate the spread of COVID-19. It may also have beneficial effects for patients who cannot tolerate longer scans and are susceptible to motion artifacts, which further degrades images. The use of lower dose may be beneficial to populations in which secondary malignancies might occur, and particularly for pediatric patients who are more sensitive to radiation and may receive many PET scans over their lifetime8.
The images included in the nuclear-medicine physician-reader study to evaluate lesion conspicuity, image quality, and diagnostic confidence consisted of the standard and LCE PET scans since low-count-only PET has previously demonstrated nondiagnostic image quality31,32,33. All three readers who analyzed the diagnostic utility of the two sets of PET scans consistently rated the low-count-enhanced scans as noninferior to the standard scans. These findings were consistent, despite the scans being acquired from PET/CT scanners of different vendor makes and models. The relative differences in ratings between the full-dose and low-count-enhanced scans were consistent across all three scanners, showing that the CNN was capable of consistently enhancing the low-count images independent of the scanner hardware and software. The intrareader repeatability on the blinded repeatability set of 10 subjects was also high, which indicated reader reliability in assessing the low-count-enhanced and standard scans. The three different readers in this study hailed from three different institutions with different PET scanner hardware and reconstruction algorithms. Consequently, intrareader variations in the assignment of DIQ and ODC scores may be reflective of their experience and comfort in interpreting different images with specific image quality created from different PET scanners.
The depiction of hypermetabolic lesions had a high sensitivity, specificity, and correlations between the two sets of images. Moreover, the combination of regression metrics such as CCC with classification metrics such as the presence/absence of lesions provided a good indication of the overall accuracy. In particular, the patient-level detection of lesions in the lymph nodes, a primary indication of cancer metastasis, was near-perfect between the low-count-enhanced images and the standard images. The prevalence rates for lesions in the muscle were low, with three out of the four lesions correctly being identified in the low-count-enhanced scans. In several instances with other subjects, the readers commonly reported diffuse regions of increased uptake as opposed to a focal uptake that could not be categorized as a lesion. Significantly, the variation observed in the lesion depiction between the standard and low-count-enhanced scans was comparable to the observed intra- and interreader variation. This suggests that the lesion conspicuity between the two scans was dependent on readers’ proclivities in addition to the images themselves, further demonstrating the noninferiority and lack of a systematic bias in the lesion-depiction capability of the low-count-enhancement technique.
While maintaining image quality is vital, PET is a quantitative technique, and thus it is essential to maintain SUV accuracy. We found high concordance for SUVmean in the aortic blood pool, liver, and right gluteus muscle—organs typically used as internal references. Similarly, SUVmax was comparable for hypermetabolic lesions also. Maintaining comparable SUV quantitation between the low-count-enhanced and standard PET images is important to assess tumor avidity and response to therapy. Coupled with the high concordance in the depiction of lesions and the SUV quantification between the standard and low-count-enhanced scans, the proposed method maintained pixelwise accuracy and imagewise quality, two criteria important for prospective image-acquisition studies34. Overall, this shows that the proposed method can be used for accurate staging and prognostication.
The study has the following limitations. During the reader study for determining hypermetabolic lesions, the three blinded readers only indicated the organs of involvement but did not indicate specific lesions. Correlating the diagnostic performance of the low-count-enhanced images as a function of lesion, VOI volume may be beneficial for staging and prognostication. Additionally, the subjects that were included in this study underwent a whole-body PET/CT protocol. The performance of the deep-learning model for other types of studies (such as for neuroimaging) was not performed. Moreover, this study only included patients scanned with FDG, while this is by far the largest clinical use case for PET-CT; the use of the deep-learning enhancement with non-FDG radiotracers may have different performance dependent upon signal-to-noise ratios and the uptake dynamics and locations. Ethnicity information was also not collected in this initial proof-of-concept study; however, future work will be necessary to determine whether there exist systematic ethnicity-based biases in model performance.
In conclusion, we have evaluated the use of a deep-learning technique to enhance image quality of fourfold count-reduced PET images in a multicenter and multivendor study. The low-count-enhanced PET scans maintained image quality and SUV accuracy as assessed by nuclear-medicine physicians from three different institutions. The testing datasets used in this study included PET images from three institutions and three different PET-CT scanners, previously unseen by the CNN, demonstrating generalizability in an external validation cohort. The variations in lesion depiction between the standard and low-count-enhanced scans were lower than intra- and interreader variation. Thus, the proposed low-count-enhancement technique is promising to enable lower radiotracer dose and to improve the efficiency of diagnostic PET imaging.
Methods
Patient population
In total, 50 subjects from three separate hospitals (20 from Institution A, 10 from Institution B, and 20 from Institution C) referred for a whole-body FDG PET/CT examination (between September 2018 and April 2019) were included in this prospective study. All subjects were scanned with Institutional Review Board approval, informed consent, and Health Insurance Portability and Accountability Act compliance. IRB approvals were obtained from the University of Southern California and Oregon Health Sciences University, while a research ethics board (REB) approval was obtained from the University of Toronto. Consecutive adult, nonpregnant patients undergoing a standard skull-base to mid-thigh FDG PET-CT were eligible. The clinical indications for the studies were for cancer diagnoses, with full demographic information presented in Table 4. In total, 26 male patients (mean age: 58 ± 17 years, range:19-90 years) and 24 female patients (mean age: 58 ± 17 years, range: 26–85 years) undergoing PET imaging with a mean FDG dose of 12.0 ± 1.9 mCi were evaluated in this study.
Low-count PET enhancement
The FDG dose and the uptake time for the PET scans was based on the standard protocol at each institution (Siemens Biograph64 Truepoint with 3 min/bed acquisition at Institution A, Siemens Biograph mCT with 2 min/bed acquisition at Institution B, and GE Discovery MI with 3 mins/bed at Institution C). The details of the reconstructions, which did not vary between the standard and low-count scans, are shown in Table 4. Low-count scans were either captured from a separate scan with fourfold shorter bed durations performed immediately following the standard scan (Institution A) or created by reconstructing shorter bed durations from the standard scan using list-mode data (Institutions B and C), which have shown image-quality equivalence previously31,35. An FDA-cleared, commercially available software product (SubtlePET, Subtle Medical, Menlo Park, CA) was used to enhance the low-count scans. This software uses a 2.5D encoder–decoder U-Net deep convolutional neural network (CNN) to perform denoising, and was trained on pairs of low- and high-count PET studies36. None of the subjects nor the institutions in this study contributed to the training of the deep network, making this a true external validation test.
Reader study for low-count-enhanced scans
Previous studies have demonstrated considerably reduced diagnostic conspicuity for hypermetabolic lesions and biased SUV measurements for fourfold low-count PET31,32,33. Consequently, the goal of this reader study was to investigate whether the low-count-enhanced PET scans using deep learning were noninferior to the standard of care. Moreover, such an approach would also lower reader fatigue and case-memorization effects that may arise if readers assessed the standard of care scans, low-count scans, and low-count-enhanced scans.
For the reader study, three board-certified nuclear-medicine physicians from three separate academic institutions evaluated the efficacy of the low-count-enhanced PET scans (G.D., E.M., and J.H. with 7, 12, and 22 years of experience). The readers evaluated two sets of PET scans per patient. One set consisted of the original 100% full-count PET scans (“standard”) and one set consisted of 25% low-count PET scans enhanced using the CNN (“low-count-enhanced”). The readers were blinded to the scan type (standard or low-count-enhanced). In addition, 10 patients were randomly chosen from the 50 original patients (4, 3, and 3 from each institution) and represented to the readers for a duplicate read to evaluate intrareader repeatability. None of the readers were made aware of this subject repetition. Overall, the 3 readers read 120 scans each (60 standard scans and 60 low-count-enhanced scans), which led to a total of 360 individual assessments.
All readers viewed the PET scans and the corresponding CT series using MIM Encore (MIM Software Inc., Cleveland, Ohio) to replicate their standard clinical reading environment. For each subject, the readers were allowed to generate multiplanar reformats and maximum-intensity projections (MIPs) as desired. After reviewing all images in the PET series, the readers were asked to score the diagnostic image quality (DIQ) on a 5-point Likert scale (1 = nondiagnostic, 2 = poor, 3 = acceptable, 4 = good, 5 = excellent image quality). The readers were also asked to indicate the number of hypermetabolic lesions depicted in the PET scan in the following organs: brain, lymph nodes, lung, liver, spleen, bone, and muscle. If more than five lesions were found in any region, the readers were instructed to report 5+ lesions. The readers were also asked to provide their overall diagnostic confidence (ODC) in interpreting the images on a Likert scale of 1–5 (1 = none, 2 = poor, 3 = acceptable, 4 = good, 5 = excellent diagnostic confidence).
A separate fourth reader (S.S., a board-certified nuclear-medicine physician with 13 years of experience) separately reviewed the standard PET scans from the 50 subjects. This reader drew volumes of interest (VOI) in the several reference regions (aortic blood pool, liver, and right gluteus muscle) and up to five abnormal lesions. For the liver, a 3-cm-diameter VOI was placed in segment VII. VOIs for up to five lesions were placed on a mix of higher- and lower-uptake lesions, to account for those with lower SUVmax. All VOIs were subsequently copied from the standard scans to the low-count-enhanced scans. The SUVmean of the VOIs was evaluated for the aortic blood pool, liver, and the right gluteus muscle as internal organ reference regions37,38,39, while the SUVmax was evaluated for the hypermetabolic lesions on the standard and low-count-enhanced scans.
Statistical analysis
DIQ and ODC for both the low-count-enhanced and standard images were analyzed using a general linear model taking into account the following sources of variation: the two different imaging methods (within subjects), the three scanners, the three readers, 50 subjects, and interaction between the imaging method and scanners. We sought to investigate both the possibility of scanner variability and the presence of heterogeneity in the difference between both imaging methods due to an interaction with the underlying scanner model. The residual standard deviation (root mean square error), with its degrees of freedom, was used as an estimate of the variability of the scores. The model for noninferiority was chosen once the sources of noise were discarded. It was used to compute 95% confidence intervals for the difference between the means of the standard and low-count-enhanced methods, and the test of noninferiority, using a noninferiority margin of 0.5 points on the DIQ and ODC Likert scales. The noninferiority margin was chosen as a midpoint between two consecutive points on the 1–5 Likert scale and using prior qualitative reader studies for PET image quality assessment40. For a significance level of 5% and a power of 90%, a noninferiority limit of 0.5 for a sample size of 50 patients would allow the standard deviation of the DIQ and ODC score differences between methods to be as large as 0.8529.
Interreader variation for DIQ metrics was assessed using an interclass correlation coefficient (ICC). Statistical significance for intrareader variation between the first and second reads was tested using Wilcoxon signed-rank tests for comparing the paired DIQ, ODC, and number of hypermetabolic lesions values in the 10 repeated cases. The number of instances where the low-count-enhanced scans did not meet a DIQ or ODC of three or higher (clinically acceptable) were tallied.
Variations between the number of hypermetabolic lesions detected by the readers between the sets of scans were evaluated using Mann–Whitney U-tests and quantified using concordance correlation coefficients (CCC). For quantitative analysis, any scans marked with 5+ lesions were treated as a scan with six lesions. A homogeneity test was performed using contingency tables using the low-count-enhanced and standard scans to assess for any systematic bias to up-stage or down-stage patients based on lesion count.
The sensitivity and specificity of the low-count-enhanced PET scans to depict lesions was evaluated at the patient level, with respect to the lesions depicted in the standard scans. The number of lesions depicted by the readers was set at a clinically relevant threshold value of one (one class for zero lesions and another class for 1+ lesions). To minimize the noise introduced by intrareader and inter-reader variation, a lesion in a subject was considered positive if it was detected by a majority of the readers (2 out of 3). To compare the variability of lesion detection between the standard and low-count-enhanced scans versus the inherent repeatability of the readers themselves between successive reads for the same patients, Cohen’s kappa was computed to compare interscan agreement (standard versus low-count-enhanced) and intrareader agreement between reads (reading 1 versus 2 for the duplicated cases). Additionally, interreader agreement and Cohen’s kappa was also computed between lesions depicted by pairs of readers (reader pairs consisting of readers 1 and 2, readers 1 and 3, and readers 2 and 3)
To evaluate equivalence for quantitative SUV between the standard and low-count-enhanced scans, the CCC and Bland–Altman plots were generated for comparing SUVmax and SUVmean for lesions and reference regions, respectively. Additionally, Mann–Whitney U-tests were used to compare systematic biases between the two scans. Finally, sensitivity and specificity analysis was performed between the standard and low-count-enhanced scans for identifying lesions above or below an SUVmax threshold of 2.541.
Overall, measures of noninferiority of image quality for the low-count-enhanced images as assessed by the DIQ and ODC metrics were the primary outcomes of this study. The assessment of nonstatistically significant SUV and lesion-detection metrics were the secondary outcomes of this study. All statistical analysis was performed using Python (version 3.6.7) using the NumPy (version 1.16) and SciPy (version 1.3) libraries. All plotting of figures’ graphical data was performed using the Python matplotlib (version 3.1) and seaborn (version 0.8.1) libraries. All statistical significance levels were set to an α=0.05 with a Holm–Bonferroni correction to adjust for multiple comparisons between multiple readers or scanner types, where necessary.
Reporting summary
Further information on research design is available in the Nature Research Reporting Summary linked to this article.
Data availability
The data from this study are not publicly available in accordance to institutional requirements governing human subject privacy considerations. The data may be made available from the authors upon reasonable request subject to permission and approval from the corresponding organizations and institutional review boards.
Code availability
The code may be made available from the authors upon reasonable request subject to permission and approval from the corresponding organizations and institutional review boards.
Change history
14 September 2021
A Correction to this paper has been published: https://doi.org/10.1038/s41746-021-00512-6
References
Gallamini, A., Zwarthoed, C. & Borra, A. Positron emission tomography (PET) in oncology. Cancers 6, 1821–1889 (2014).
Zasadny, K. R. & Wahl, R. L. Standardized uptake values of normal tissues at PET with 2-[fluorine-18]-fluoro-2-deoxy-D-glucose: variations with body weight and a method for correction. Radiology 189, 847–950 (1993).
Strauss, L. & Conti, P. S. The applications of PET in clinical oncology. J. Nucl. Med. 32, 623–650 (1991).
Thie, J. A. Understanding the standardized uptake value, its methods, and implications for usage. J. Nucl. Med. 45, 1431–1434 (2004).
Chen, K. T. et al. MR-assisted PET motion correction in simultaneous PET/MRI studies of dementia subjects. J. Magn. Reson. Imaging 48, 1288–1296 (2018).
Callahan, J., Binns, D., Dunn, L. & Kron, T. Motion effects on SUV and lesion volume in 3D and 4D PET scanning. Australas. Phys. Eng. Sci. Med. 34, 489–495 (2011).
Meulepas, J. M. et al. Radiation exposure from pediatric CT scans and subsequent cancer risk in the Netherlands. J. Natl Cancer Inst. 111, 256–263 (2019).
Pearce, M. S. et al. Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 380, 499–505 (2012).
Roberts, F. O. et al. Radiation dose to PET technologists and strategies to lower occupational exposure. J. Nucl. Med. Technol. 33, 44–47 (2005).
Huang, B., Law, M. W. M. & Khong, P. L. Whole-body PET/CT scanning: estimation of radiation dose and cancer risk. Radiology 251, 166–174 (2009).
Ardila, D. et al. End-to-end lung cancer screening with three-dimensional deep learning on low-dose chest computed tomography. Nat. Med. 25, 954–961 (2019).
McKinney, S. M. et al. International evaluation of an AI system for breast cancer screening. Nature 577, 89–94 (2020).
Campanella, G. et al. Clinical-grade computational pathology using weakly supervised deep learning on whole slide images. Nat. Med. 25, 1301–1309 (2019).
Esteva, A. et al. Dermatologist-level classification of skin cancer with deep neural networks. Nature 542, 115–118 (2017).
Chen, H. et al. Low-Dose CT with a residual encoder-decoder convolutional neural network. IEEE Trans. Med. Imaging 36, 2524–2535 (2017).
Chaudhari, A. S. et al. Super-resolution musculoskeletal MRI using deep learning. Magn. Reson. Med. 80, 2139–2154 (2018).
Kaplan, S. & Zhu, Y. Full-dose PET image estimation from low-dose PET image using deep learning: a pilot study. J. Digit. Imaging 32, 773–778 (2019).
Kang, J. et al. Prediction of standard-dose brain PET image by using MRI and low-dose brain [18F]FDG PET images. Med. Phys. 42, 5301–5309 (2015).
Chen, K. T. et al. Ultra-low-dose 18F-florbetaben amyloid PET imaging using deep learning with multi-contrast MRI inputs. Radiology 290, 649–656 (2019).
Wang, Y. et al. 3D conditional generative adversarial networks for high-quality PET image estimation at low dose. Neuroimage 174, 550–562 (2018).
Wang, Y. R. (Joyce) et al. Artificial intelligence enables whole-body positron emission tomography scans with minimalradiation exposure. Eur. J. Nucl. Med. Mol. Imaging Aug; 48:2771-2781. (2021).
Recht M. P. et al. Using Deep Learning to Accelerate Knee MRI at 3 T: Results of an Interchangeability Study. AJR Am J Roentgenol 215:1421-1429 (2020).
Chaudhari, A. S. et al. Diagnostic accuracy of quantitative multi-contrast 5-minute knee MRI using prospective artificial intelligence image quality enhancement title. Am. J. Roentgenol. 216, 1614–1625 (2021).
Akamatsu, G. et al. Improvement in PET/CT image quality with a combination of point-spread function and time-of-flight in relation to reconstruction parameters. J. Nucl. Med. 53, 1716–1722 (2012).
Neyshabur, B., Bhojanapalli, S., McAllester, D. & Srebro, N. Exploring generalization in deep learning. Adv. Neural Inf. Process. Syst. 2017, 5948–5957 (2017).
Zech, J. R. et al. Variable generalization performance of a deep learning model to detect pneumonia in chest radiographs: a cross-sectional study. PLoS Med. 15, 1–17 (2018).
Paschali, M., Conjeti, S., Navarro, F. & Navab, N. Generalizability vs. robustness: investigating medical imaging networks using adversarial examples. Med. Image Comput. Comput. Assist. Interv. https://doi.org/10.1007/978-3-030-00928-1_56 (2018).
Bluemke, D. A. et al. assessing radiology research on artificial intelligence: a brief guide for authors, reviewers, and readers—from the radiology editorial board. Radiology. 294, 487–489 (2019).
Julious, S. A. Sample sizes for clinical trials with Normal data. Stat. Med. 23, 1921–1986 (2004).
Sekine, T. et al. Reduction of 18 F-FDG dose in clinical PET/MR imaging by using silicon photomultiplier detectors. Radiology 286, 249–259 (2018).
Gatidis, S. et al. Towards tracer dose reduction in PET studies: simulation of dose reduction by retrospective randomized undersampling of list-mode data. Hell. J. Nucl. Med. 19, 15–18 (2016).
Seith, F. et al. Simulation of tracer dose reduction in 18F-FDG PET/MRI: effects on oncologic reading, image quality, and artifacts. J. Nucl. Med. 58, 1699–1705 (2017).
Schaefferkoetter, J. D. et al. Quantitative accuracy and lesion detectability of low-dose 18F-FDG PET for lung cancer screening. J. Nucl. Med. 58, 399–405 (2017).
Chaudhari, A. S. et al. Prospective deployment of deep learning in MRI: a framework for important considerations, challenges, and recommendations for best practices. J. Magn. Reson. Imaging 54, 357–371 (2021).
Chen, K. T. et al. True ultra-low-dose amyloid PET/MRI enhanced with deep learning for clinical interpretation. Eur. J. Nucl. Med. Mol. Imaging. 48, 2416–2425. (2021).
Ronneberger, O., Fischer, P. & Brox, T. U-Net: convolutional networks for biomedical image segmentation. Med. Image Comput. Comput. Assist. Interv. Lecture Notes in Computer Science, vol 9351. https://doi.org/10.1007/978-3-319-24574-4_28 (2015).
Gholami, S., Salavati, A., Houshmand, S., Werner, T. J. & Alavi, A. Assessment of atherosclerosis in large vessel walls: a comprehensive review of FDG-PET/CT image acquisition protocols and methods for uptake quantification. J. Nucl. Cardiol. 22, 468–479 (2015).
Wahl, R. L., Jacene, H., Kasamon, Y. & Lodge, M. A. From RECIST to PERCIST: evolving considerations for PET response criteria in solid tumors. J. Nucl. Med. 50, 122–151 (2009).
Ahlman, M. A. et al. Internal tissue references for 18Fluorodeoxyglucose vascular inflammation imaging: Implications for cardiovascular risk stratification and clinical trials. PLoS ONE 12, 1–12 (2017).
Paldino, M. J. et al. Comparison of the diagnostic accuracy of PET/MRI to PET/CT-acquired FDG brain exams for seizure focus detection: a prospective study. Pediatr. Radiol. 47, 1500–1507 (2017).
Nestle, U. et al. Comparison of different methods for delineation of 18F-FDG PET-positive tissue for target volume definition in radiotherapy of patients with non-Small cell lung cancer. J. Nucl. Med. 46, 1342–1348 (2005).
Acknowledgements
We would like to acknowledge Dr. Marta Garcia-Granero Márquez for providing assistance with the statistical analysis in this paper.
Author information
Authors and Affiliations
Contributions
All authors meet the criteria detailed in Author Instructions. Study conception and design: AC, EM, GD, PG, GZ, HJ. Acquisition of Data: PG, HG, AB, and TZ. Interpretation of imaging data: EM, GD, SS, and HJ. Creation of figures and tables: AC, EM, GD, PG, EG, and GZ. Analysis of data: AC, EM, GD TZ, EG, and GZ. Primary drafting of paper and figures: AC, EM, and GD. Critical revision of paper: All authors. Final approval of the paper: All authors. Equal contribution: AC, EM, and GD.
Corresponding author
Ethics declarations
Competing interests
A.C. is affiliated with Stanford University and Subtle Medical. This work was fully conducted under a consulting relationship for Subtle Medical where A.C. is a shareholder. Activities not related to the present article: consulting services to Skope MR, Culvert Engineering, Edge Analytics, Image Analysis Group, ICM, and Chondrometrics GmbH; and shareholder in LVIS Corp. and Brain Key. E.M. has provided consulting services to Ipsen, Curium, and AAA. G.D. has provided consulting services to Genentech Inc. and receives research support from Kheiron Medical Technologies Inc., Dimensional Mechanics Inc., and GE Healthcare. P.G., H.G., and T.Z. are employees and shareholders of Subtle Medical. A.B. has no disclosures to report. S.S. has provided consulting services to Subtle Medical. Activities not related to the present article: disclosed employment as chief medical officer for approximately one year with Sirtex Medical Inc. E.G. is a cofounder and shareholder of Subtle Medical. G.Z. is a cofounder and shareholder of Subtle Medical. Activities not related to the present article: institutional grant from GE Healthcare. H.J. has received research support from Subtle Medical. Intellectual property: E.G. and G.Z. are coinventors of the following patent “Dose Reduction For Medical Imaging Using Deep Convolutional Neural Networks”, with specific details below: Application Number: PCT/US2018/029103 Publication Number: WO/2018/200493 Applicant: The Board of Trustees of The Leland Stanford Junior University Inventors: Greg Zaharchuk, John M. Pauly and Enhao Gong Aspect Covered: The technology described in this patent was used to perform the low-dose PET-image enhancement described in this paper.
Additional information
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Chaudhari, A.S., Mittra, E., Davidzon, G.A. et al. Low-count whole-body PET with deep learning in a multicenter and externally validated study. npj Digit. Med. 4, 127 (2021). https://doi.org/10.1038/s41746-021-00497-2
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41746-021-00497-2
This article is cited by
-
Deep learning-based techniques for estimating high-quality full-dose positron emission tomography images from low-dose scans: a systematic review
BMC Medical Imaging (2024)
-
Comparison of the Accuracy of a Deep Learning Method for Lesion Detection in PET/CT and PET/MRI Images
Molecular Imaging and Biology (2024)
-
Denoising approaches by SubtlePET™ artificial intelligence in positron emission tomography (PET) for clinical routine application
Clinical and Translational Imaging (2024)
-
Intelligent ultrafast total-body PET for sedation-free pediatric [18F]FDG imaging
European Journal of Nuclear Medicine and Molecular Imaging (2024)
-
Deep learning-based PET image denoising and reconstruction: a review
Radiological Physics and Technology (2024)