Abstract
Using an amorphous silicon (a-Si) EPID for transit dosimetry requires detailed characterization of its dosimetric response in a variety of conditions. In this study, a measurement-based model was developed to calibrate an a-Si EPID response to dose for transit dosimetry by comparison with a reference ionization chamber. The ionization chamber reference depth and the required additional buildup thickness for electronic portal imaging devices (EPID) transit dosimetry were determined. The combined effects of changes in radiation field size, phantom thickness, and the off-axis distance on EPID transit dosimetry were characterized. The effect of scattered radiation on out-of-field response was investigated for different field sizes and phantom thicknesses by evaluation of the differences in image profiles and in-water measured profiles. An algorithm was developed to automatically apply these corrections to EPID images based on the user-specified field size and phantom thickness. The average phantom thickness and an effective field size were used for IMRT fields, and images were acquired in cine mode in the presence of an anthropomorphic phantom. The effective field size was defined as the percentage of the jaw-defined field that was involved during the delivery. Nine head and neck dynamic IMRT fields were tested by comparison with a MatriXX two-dimensional array dosimeter using the Gamma (3 %, 3 mm) evaluation. A depth of 1.5 cm was selected as the ionization chamber reference depth. An additional 2.2 mm of copper buildup was added to the EPID. Comparison of EPID and MatriXX dose images for the tested fields showed that using a 10 % threshold, the average number of points with Gamma index <1 was 96.5 %. The agreement in the out-of field area was shown by selection of a 2 % threshold which on average resulted in 94.8 % of points with a Gamma index <1. The suggested method is less complicated than previously reported techniques and can be used for all a-Si EPIDs regardless of the manufacturer.
Similar content being viewed by others
References
Antonuk LE (2002) Electronic portal imaging devices: a review and historical perspective of contemporary technologies and research. Phys Med Biol 47:R31–R65
Bailey DW, Kumaraswamy L, Podgorsak MB (2009) An effective correction algorithm for off-axis portal dosimetry errors. Med Phys 36(9):4089–4094
Berry SL, Polvorosa CS, Wuu C (2010) A field size specific backscatter correction algorithm for accurate EPID dosimetry. Med Phys 37(6):2425–2435
Chen J, Chuang CF, Morin O, Aubin M, Pouliot J (2006) Calibration of an amorphous silicon flat panel portal imager for exit beam dosimetry. Med Phys 33(3):584–594
Chen J, Moran JA, Roberts DA, El-Mohri Y, Antonuk LE, Fraass BA (2007) Performance of a direct-detection active matrix flat panel dosimeter (AMFPD) for IMRT measurements. Med Phys 34(12):4911–4922
Childress NL, Dong L, Rosen II (2002) Rapid radiographic film calibration for IMRT verification using automated MLC fields. Med Phys 29(10):2384–2390
Chung H, Li J, Samant S (2011) Feasibility of using two-dimensional array dosimeter for in vivo dose reconstruction via transit dosimetry. J Appl Clin Med Phys 12(3):90–111
Chytyk K, McCurdy BM (2009) Comprehensive fluence model for absolute portal dose image prediction. Med Phys 36(4):1389–1398
Greer PG (2005) Correction of pixel sensitivity variation and off-axis response for amorphous silicon EPID dosimetry. Med Phys 32(12):3558–3568
Greer PG, Popescu CC (2003) Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapy. Med Phys 30:1618–1627
Greer P, Vial P (2011) EPID Dosimetry. AIP Conf Proc 1345:129–144
Grein EE, Lee R, Luchka K (2002) An investigation of a new amorphous silicon electronic portal imaging device for transit dosimetry. Med Phys 29(10):2262–2268
Han Z, Ng SK, Bhagwat MS, Lyatskaya Y, Zygmanski P (2010) Evaluation of MatriXX for IMRT and VMAT dose verifications in peripheral dose regions. Med Phys 37(7):3704–3714
Hansen VN, Evans PM, Swindell W (1996) The application of transit dosimetry to precision radiotherapy. Med Phys 23(5):713–721
Herman MG, Balter JM, Jaffray DA, McGee KP, Munro P, Shalev S, Van Herk M, Wong JW (2001) Clinical use of electronic portal imaging: report of AAPM radiation therapy committee task group 58. AAPM Refresher Course, Salt Lake City
Kavuma A, Glegg M, Currie G, Elliot A (2008) Assessment of dosimetrical performance in 11 Varian a-Si500 electronic portal imaging devices. Phys Med Biol 53:6893–6909
Li JG, Yan G, Liu C (2009) Comparison of two commercial detector arrays for IMRT quality assurance. J Appl Clin Med Phys 10(2):62–74
Low AD, Harms WB, Mutic S, Purdy JA (1998) A technique for the quantitative evaluation of dose distributions. Med Phys 25(5):656–661
Low DA, Moran JM, Dempsey JF, Dong L, Oldham M (2011) Dosimetry tools and techniques for IMRT. Med Phys 38(3):1313–1338
Mans A, Wendling M, McDermott LN, Sonke JJ, Tielenburg R, Vijlbrief R, Mijnheer B, van Herk M, Stroom JC (2010) Catching errors with in vivo EPID dosimetry. Med Phys 37(6):2638–2644
Mans A, Remeijer P, Olaciregui-Ruiz I, Wendling M, Sonke JJ, Minjheer B, van Herk M, Stroom JC (2010) 3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetry. Radiother Oncol 94:181–187
McCurdy BM, Greer PB (2009) Dosimetric properties of an amorphous-silicon EPID used in continuous acquisition mode for application to dynamic and arc IMRT. Med Phys 36(7):3028–3039
McCurdy BMC, Luchka KB, Pistorius S (2001) Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging device. Med Phys 28:911–924
McDermott LN, Louwe RJW, Sonke JJ, van Herk MB, Mijnheer BJ (2004) Dose-response and ghosting effects of an amorphous silicon electronic portal imaging device. Med Phys 31(2):285–295
McDermott LN, Wendling M, Snoke JJ, van Herk MB, Mijnheer BJ (2007) Replacing pre-treatment verification with in vivo EPID dosimetry for prostate IMRT. Int J Radiat Oncol Biol Phys 67:1568–1577
McNutt TR, Mackie TR, Reckwerdt P, Papanikolaou N, Paliwal BR (1996) Calculation of portal dose using the convolution/superposition method. Med Phys 23:527–535
Nijsten SMJJG, van Elmpt WJC, Jacobs M, Minjheer BJ, Dekker ALAJ, Lambin P, Minken AWH (2007) A global calibration model for a-Si EPIDs used for transit dosimetry. Med Phys 34(10):3872–3880
Olaciregui-Ruiz I, Rozendaal R, Mijnheer B, van Herk M, Mans A (2013) Automatic in vivo portal dosimetry of all treatments. Phys Med Biol 58:8253–8264
Parent L, Fielding AL, Dance DR, Seco J, Evans PM (2007) Amorphous silicon EPID calibration for dosimetric applications: comparison of a method based on Monte Carlo prediction of response with existing techniques. Phys Med Biol 52:3351–3368
Partridge M, Hesse BM, Müller L (2002) A performance comparison of direct- and indirect- detection flat panel imagers. Nucl Instrum Methods Phys Res A 484:351–363
Piermattei A, Fidanzio A, Azario L et al (2009) In patient dose reconstruction using a cine acquisition for dynamic arc radiation therapy. Med Biol Eng Comput 47:425–433
Roberson PL, Moran JM, Kulasekere R (2008) Radiographic film dosimetry for IMRT fields in the nearsurface buildup region. J Appl Clin Med Phys 9(4):87–97
Rowshanfarzad P, McCurdy BMC, Sabet M, Lee C, O’Connor DJ, Greer PB (2010) Measurement and modeling of the effect of support arm backscatter on dosimetry with a Varian EPID. Med Phys 37(5):2269–2278
Sabet M, Menk FW, Greer PB (2010) Evaluation of an a-Si EPID in direct detection configuration as a water equivalent dosimeter for transit dosimetry. Med Phys 37:1459–1467
Sabet M, Rowshanfarzad P, Vial P, Menk FW, Greer PB (2012) Transit dosimetry in IMRT with an a-Si EPID in direct detection configuration. Phys Med Biol 57:N259–N306
Sharma DS, Mhatre V, Heigrujam M, Talapatra K, Mallik S (2010) Portal dosimetry for pretreatment verification of IMRT plan: a comparison with 2D ion chamber array. J Appl Clin Med Phys 11(4):238–248
Siebers JV, Kim JO, Ko L, Keall PJ (2004) Monte Carlo computation of dosimetric amorphous silicon electronic portal images. Med Phys 31:2135–2146
Steciw S, Warkentin B, Rathee S, Fallone BG (2005) Three-dimensional IMRT verification with a flat-panel EPID. Med Phys 32:600–612
Talamonti C, Casati M, Bucciolini M (2006) Pretreatment verification of IMRT absolute dose distributions using a commercial a-Si EPID. Med Phys 33:4367–4378
Tyner E, McClean B, McCavana P, Wetterstedt S (2009) Experimental investigation of the response of an a-Si EPID to an unflattened photon beam from an Elekta Precise linear accelerator. Med Phys 36(4):1318–1329
Van Elmpt W, McDermott L, Nijsten S, Wendling M, Lambin P, Mijnheer B (2008) A literature review of electronic portal imaging for radiotherapy dosimetry. Radiother Oncol 88:289–309
Vial P, Greer PB, Oliver L, Baldock C (2008) Initial evaluation of a commercial EPID modified to a novel direct-detection configuration for radiotherapy dosimetry. Med Phys 35(10):4362–4374
Wendling M, Louwe RJ, McDermott LN, Sonke JJ, van Herk M, Mijnheer BJ (2006) Accurate two dimensional IMRT verification using a back projection EPID dosimetry method. Med Phys 33:259–273
Winkler P, Hefner A, Georg D (2005) Dose-response characteristics of an amorphous silicon EPID. Med Phys 32(10):3095–3105
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Sabet, M., Rowshanfarzad, P., Menk, F.W. et al. Transit dosimetry in dynamic IMRT with an a-Si EPID. Med Biol Eng Comput 52, 579–588 (2014). https://doi.org/10.1007/s11517-014-1161-y
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11517-014-1161-y