Abstract
The purpose of this study was to determine the application accuracy of a new frameless marker system for interactive intraoperative localization of intracranial lesions. The influence of image quality, registration error, repeatability, and marker distribution on the application accuracy were analyzed and compared. A phantom was mounted with the standard Z-D ring and also implanted with frameless marker system, which randomly distributed on the surface. The phantom was scanned as routine with 1 mm and 2 mm sections. The pixel sizes were used 1.18×1.18 and 0.59×0.59. The two systems were tested under different image quality and registration. The target point was digitized and the coordinates were recorded and compared with reference points. The difference between two systems were tested with paired t-test. Image data were loaded into a SUN Workstation and registered with NSPS.4.0 software. The coordinate of each fiducial marker was recorded into a file as the reference. The tip of each semi-invasive fiducial marker was digitized to achieve a frameless transformation matrix, and the special points on the Z-D ring were digitized to achieve a frame-based transformation matrix. The differences from the reference points were used as the deviation from “true point”. The mean square root (RMS) was calculated to show the sum of vectors. The results of 2 mm section group showed that the registration error of frame-based system is 3.42±0.22 mm and the error of the frameless system is 1.01±0.63 mm (P<0.001). The RMS are 2.57±0.54 mm and 1.53±0.65 mm respectively (P<0.001). The RMS of error registration (one point off 5 mm) are 5.01±0.26 mm and 2.23±0.13 mm respective (P=0.003). The results of 1mm section group showed that the RMS are 1.20±0.42 mm and 0.90±0.47 mm respectively (P=0.121). The higher the quality (the thinner scan thickness) of image it is, the better the application accuracy will be (P=0.001 and 0.032 respectively). These preliminary results showed that the frameless semi-invasive fiducial marker system can provide clinical acceptable accurate localization as the frame based surgical localization system did. There is no significant difference between the experimental and clinical results. The higher the quality of image it is, the better the application accuracy will be. But there is no significant difference between 1mm sections and 3 mm sections of MRI images.
Chapter PDF
Key words
References
Bucholz RD, Smith KR (1993) A comparison of sonic digitizers versus light emitting diode-based location. In Maciunas RJ (ed) Interactive Image-Guided Neurosurgery. AANS Publications Committee, Park Ridge 1993. Pp 179–200
Bucholz RD, Ho HW, and Rubin JP.(1993) Variables affecting the accuracy of stereotactic localization using computerized tomography. J Neurosurg 79: 667–673
Golfinos JC, Fitzpatrick BC, Smith LR, Spetzler R (1995) Clinical use of a frameless stereotactic arm: Results of 325 cases. J Neurosurg 83: 197–205
Heilbrun MP, McDonald P, Wiker C, et al (1992) Stereotactic localization and guidance using a machine vision technique. Stereotact Func Neurosurg 58: 94–98
Murphy MA, Barnett GH, Kormos DW, Waisemberger J (1993) Astrocytoma resection using an interactive frameless stereotactic wand: An early experience. J Clin Neurosci 1: 3337
Kato A, Yoshimine T, Hayakawa T, et al (1991) A frameless, armless navigational system for computer-assisted neurosurgery. J Neurosurg 74: 845–849
Li Q.H., Holdener H.J., Zamorano L., King P., Jiang Z.W., and Diaz F.: Computer assisted insertion of pedicle screws. Lecture Notes in Computer Science 1131. Eds by Karl Heinz H_hne and Ron Kikinis, page 571–581, 1996.
Maciunas RJ, Fitzpatrick JM, Galloway RL, Allen GS. Beyond stereotaxy: Extreme levels of application accuracy are provided by implantable fiducial markers for interactive image-guided neurosurgery. In Maciunas RJ (ed) Interactive Imaged-Guided Surgery. AANS Publications Committee, Park Ridge 1993, pp 259–270
Nolte LP, Zamorano L, Jiang Z, Wang Q, Langlotz F, Berlemann U (1995) Image-guided insertion of transpedicular screws: A laboratory set-up. Spine 20: 497–500
Takizawa T (1993) Neurosurgical navigation using a noninvasive stereoadapter. Surg Neurol 40: 1–7
Watanabe E, Mayanagi Y, Kosugi Y, Manaka S, Takakura K (1991) Open surgery assisted by the neuronavigator, a stereotactic, articulated, sensitive arm. Neurosurg 28: 792–800
Zamorano L, Nolte L, Jiang C, Kadi M (1993) Image-guided neurosurgery: Frame based versus ftameless approaches. Neurosurgical Operative Atlas 3: 402–422
Zamorano L, Nolte L, Kadi M, Jiang Z (1993) Interactive intra-operative localization using an infraredbased system. Neuro Res 15: 290–298
Zamorano L. Nolte LP, Kadi AM, Jiang Z (1994) Interactive intraoperative localization using an infraredbased system. Stereotact Funct Neurosurg 63: 84–88
Zamorano L, Kadi M, Jiang Z, Diaz FG (1994) Zamorano-Dujovny multipurpose neurosurgical imageguided localizing unit: Experience in 866 consecutive cases of “open stereotaxis”. Stereotact Funct Neurosurg 63: 45–51
Author information
Authors and Affiliations
Editor information
Rights and permissions
Copyright information
© 1998 Springer-Verlag Berlin Heidelberg
About this paper
Cite this paper
Li, Q., Zamorano, L., Jiang, Z., Vinas, F., Diaz, F. (1998). The application accuracy of the frameless implantable marker system and analysis of related affecting factors. In: Wells, W.M., Colchester, A., Delp, S. (eds) Medical Image Computing and Computer-Assisted Intervention — MICCAI’98. MICCAI 1998. Lecture Notes in Computer Science, vol 1496. Springer, Berlin, Heidelberg. https://doi.org/10.1007/BFb0056208
Download citation
DOI: https://doi.org/10.1007/BFb0056208
Published:
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-540-65136-9
Online ISBN: 978-3-540-49563-5
eBook Packages: Springer Book Archive